Accuracy & Precision
Accuracy and precision data obtained under optimized conditions may not be possible to reproduce in your application. The data below were obtained under normal QPCR set-up conditions. On the other hand, high precision data indicates that the liquid handling is stable. Market leading precision can easily be reproduced at your lab. High precision and accuracy also grants that dispense performance is very robust. Changes in viscosity will not have the same impact on dispensed volume seen in competing systems. So the dispense data below is what you can expect from Rob™ in reality.
50 ul pipette tip with filter
Volume (µl) |
Mean (µl) |
Precision (C.V.) |
1 |
1.00 |
3.78 |
2 |
1.99 |
1.83 |
5 |
5.00 |
0.76 |
10 |
10.02 |
0.38 |
Ten single dispenses in ten PCR tubes (0.2 ml).
50 ul pipette tip with filter, OneDip™
Volume (µl) |
Mean (µl) |
Precision (C.V.) |
2 |
2.03 |
1.93 |
5 |
5.05 |
1.67 |
Three series dispenses in 3 x3 PCR tubes. Tubes were pre-filled with 15 µl master mix to mimic a QPCR set-up.